PV-CT-WALL-001

Software PASS
Units:
Account

Software verification and independent professional review are tracked separately. This page never claims professional approval — only a qualified reviewer, using their own judgement, credentials, and the reviewer packet below, can determine whether a case is acceptable.

CT adjacent wall

v1.0.0 · ct
Purpose
CT wall scatter point and barrier set contribution.
Required reviewer qualification
Diagnostic medical physicist or qualified expert
Formula IDs
CT.SCATTER.POINT, CT.BARRIER.SET
Dataset binding
METADATA_CITATION_ONLY

Assumptions

  • Caller-supplied reference scatter

Applicability limits

  • CT adjacent occupied walls

Known limitations

  • Vertical adjacency incomplete warning expected for single-direction set

Independent calculation instructions: Distance-correct CT scatter and apply wall transmission.

Software results

Compared against a locked regression baseline. Comparison state: PASS · Verification state: SOFTWARE_VERIFICATION_PASS

Software result values
FieldValue
directionwall
weeklyMgy0.002222
meetsGoaltrue

Input hash: 8cb7ca05f7081592 · Result hash: 01448cc823b1b495

Reviewer packet

Full markdown packet a qualified reviewer uses to perform an independent calculation.

# Reviewer packet — PV-CT-WALL-001

**Title:** CT adjacent wall
**Version:** 1.0.0
**Modality:** ct
**Software state:** SOFTWARE_VERIFICATION_PASS
**Independent review status:** READY_FOR_INDEPENDENT_REVIEW (pending human entry)

## Boundaries

- Software verification is not professional validation.
- Self-review does not qualify as independent review.
- Do not fabricate independent results.
- Licensed package citation: `ncrp147-diagnostic-uat-0.3.0` (METADATA_CITATION_ONLY).
- Numeric software lock uses demonstration transmission parameters only.

## Purpose

CT wall scatter point and barrier set contribution.

## Formula IDs

- `CT.SCATTER.POINT`
- `CT.BARRIER.SET`

## Independent calculation instructions

Distance-correct CT scatter and apply wall transmission.

## Required reviewer qualification

Diagnostic medical physicist or qualified expert

## Software results (comparison values; not pre-rounded away)

```json
{
  "direction": "wall",
  "weeklyMgy": 0.0022222222222222222,
  "meetsGoal": true
}
```

## Display rounding (informational only)

```json
{
  "direction": "wall",
  "weeklyMgy": "0.002222",
  "meetsGoal": "true"
}
```

## Hashes

- Input hash: `8cb7ca05f7081592`
- Result hash: `01448cc823b1b495`

## Assumptions

- Caller-supplied reference scatter

## Known limitations

- Vertical adjacency incomplete warning expected for single-direction set

Independent result entry

Demo-only entry for this CONTROLLED_UAT session. Values are held in server memory (not a database) and reset when the server restarts.

SYNTHETIC / DEMO ENTRY — not a real independent professional review

Independently computed values

Discrepancies

No discrepancies opened for this case.

Professional approval attempt

Approval is blocked while any discrepancy is open, while the case has not passed software verification, when no independent result has been recorded, or when credential/authority requirements are not met. Any approval recorded in this environment is a CONTROLLED UAT ELECTRONIC APPROVAL RECORD — not a qualified digital signature — and does not itself constitute professional approval.