NONPRODUCTION / CONTROLLED UAT · PROFESSIONAL VALIDATION REQUIRED

PV-RAD-FLOOR-001

Software PASS
Units:
Account

Software verification and independent professional review are tracked separately. This page never claims professional approval — only a qualified reviewer, using their own judgement, credentials, and the reviewer packet below, can determine whether a case is acceptable.

Floor-below receptor

v1.0.0 · radiography
Purpose
Vertical floor adjacency contribution.
Required reviewer qualification
Diagnostic medical physicist or qualified expert
Formula IDs
RAD.FLOOR_CEILING.SCATTER
Dataset binding
METADATA_CITATION_ONLY

Assumptions

  • Synthetic concrete transmission params

Applicability limits

  • Floor below occupied spaces

Known limitations

  • Not a sealed structural calculation

Independent calculation instructions: Apply floor transmission to unshielded weekly contribution.

Software results

Compared against a locked regression baseline. Comparison state: PASS · Verification state: SOFTWARE_VERIFICATION_PASS

Software result values
FieldValue
boundaryfloor
weeklyDoseMgy0.000000
transmission0.000000

Input hash: 4061fd8e097767b1 · Result hash: ed62541307c55f54

Reviewer packet

Full markdown packet a qualified reviewer uses to perform an independent calculation.

# Reviewer packet — PV-RAD-FLOOR-001

**Title:** Floor-below receptor
**Version:** 1.0.0
**Modality:** radiography
**Software state:** SOFTWARE_VERIFICATION_PASS
**Independent review status:** READY_FOR_INDEPENDENT_REVIEW (pending human entry)

## Boundaries

- Software verification is not professional validation.
- Self-review does not qualify as independent review.
- Do not fabricate independent results.
- Licensed package citation: `ncrp147-diagnostic-uat-0.3.0` (METADATA_CITATION_ONLY).
- Numeric software lock uses demonstration transmission parameters only.

## Purpose

Vertical floor adjacency contribution.

## Formula IDs

- `RAD.FLOOR_CEILING.SCATTER`

## Independent calculation instructions

Apply floor transmission to unshielded weekly contribution.

## Required reviewer qualification

Diagnostic medical physicist or qualified expert

## Software results (comparison values; not pre-rounded away)

```json
{
  "boundary": "floor",
  "weeklyDoseMgy": 6.640814265914661e-9,
  "transmission": 2.6563257063658646e-8
}
```

## Display rounding (informational only)

```json
{
  "boundary": "floor",
  "weeklyDoseMgy": "0.000000",
  "transmission": "0.000000"
}
```

## Hashes

- Input hash: `4061fd8e097767b1`
- Result hash: `ed62541307c55f54`

## Assumptions

- Synthetic concrete transmission params

## Known limitations

- Not a sealed structural calculation

Independent result entry

Demo-only entry for this CONTROLLED_UAT session. Values are held in server memory (not a database) and reset when the server restarts.

SYNTHETIC / DEMO ENTRY — not a real independent professional review

Independently computed values

Discrepancies

No discrepancies opened for this case.

Professional approval attempt

Approval is blocked while any discrepancy is open, while the case has not passed software verification, when no independent result has been recorded, or when credential/authority requirements are not met. Any approval recorded in this environment is a CONTROLLED UAT ELECTRONIC APPROVAL RECORD — not a qualified digital signature — and does not itself constitute professional approval.