Software verification and independent professional review are tracked separately. This page never claims professional approval — only a qualified reviewer, using their own judgement, credentials, and the reviewer packet below, can determine whether a case is acceptable.
Multiple primary beam directions
v1.0.0 · radiography- Purpose
- Aggregate two primary directions and identify the dominant contributor.
- Required reviewer qualification
- Diagnostic medical physicist or qualified expert
- Formula IDs
- RAD.PRIMARY.REQUIRED_TRANSMISSION
- Dataset binding
- METADATA_CITATION_ONLY
Assumptions
- Two independent primary directions
Applicability limits
- Multi-direction primary receptors
Known limitations
- Demonstration transmission params
Independent calculation instructions: Sum weekly doses for both directions and compare to goal.
Software results
Compared against a locked regression baseline. Comparison state: PASS · Verification state: SOFTWARE_VERIFICATION_PASS
| Field | Value |
|---|---|
| directionCount | 2.000000 |
| combinedWeeklyMgy | 0.007713 |
| meetsGoal | true |
| dominantDirectionId | wall-b |
Input hash: 83df0e2a05d22557 · Result hash: d311019d551deade
Reviewer packet
Full markdown packet a qualified reviewer uses to perform an independent calculation.
# Reviewer packet — PV-RAD-PRI-002
**Title:** Multiple primary beam directions
**Version:** 1.0.0
**Modality:** radiography
**Software state:** SOFTWARE_VERIFICATION_PASS
**Independent review status:** READY_FOR_INDEPENDENT_REVIEW (pending human entry)
## Boundaries
- Software verification is not professional validation.
- Self-review does not qualify as independent review.
- Do not fabricate independent results.
- Licensed package citation: `ncrp147-diagnostic-uat-0.3.0` (METADATA_CITATION_ONLY).
- Numeric software lock uses demonstration transmission parameters only.
## Purpose
Aggregate two primary directions and identify the dominant contributor.
## Formula IDs
- `RAD.PRIMARY.REQUIRED_TRANSMISSION`
## Independent calculation instructions
Sum weekly doses for both directions and compare to goal.
## Required reviewer qualification
Diagnostic medical physicist or qualified expert
## Software results (comparison values; not pre-rounded away)
```json
{
"directionCount": 2,
"combinedWeeklyMgy": 0.007712533688181307,
"meetsGoal": true,
"dominantDirectionId": "wall-b"
}
```
## Display rounding (informational only)
```json
{
"directionCount": "2.000000",
"combinedWeeklyMgy": "0.007713",
"meetsGoal": "true",
"dominantDirectionId": "wall-b"
}
```
## Hashes
- Input hash: `83df0e2a05d22557`
- Result hash: `d311019d551deade`
## Assumptions
- Two independent primary directions
## Known limitations
- Demonstration transmission params
Independent result entry
Demo-only entry for this CONTROLLED_UAT session. Values are held in server memory (not a database) and reset when the server restarts.
Discrepancies
No discrepancies opened for this case.
Professional approval attempt
Approval is blocked while any discrepancy is open, while the case has not passed software verification, when no independent result has been recorded, or when credential/authority requirements are not met. Any approval recorded in this environment is a CONTROLLED UAT ELECTRONIC APPROVAL RECORD — not a qualified digital signature — and does not itself constitute professional approval.