NONPRODUCTION / CONTROLLED UAT · PROFESSIONAL VALIDATION REQUIRED

PV-RAD-SRC-001

Software PASS
Units:
Account

Software verification and independent professional review are tracked separately. This page never claims professional approval — only a qualified reviewer, using their own judgement, credentials, and the reviewer packet below, can determine whether a case is acceptable.

Multiple source positions

v1.0.0 · radiography
Purpose
Aggregate multi-source weekly contributions.
Required reviewer qualification
Diagnostic medical physicist or qualified expert
Formula IDs
RAD.MULTI_SOURCE.AGGREGATE
Dataset binding
METADATA_CITATION_ONLY

Assumptions

  • Precomputed per-source weekly doses

Applicability limits

  • Rooms with multiple tubes/positions

Known limitations

  • Per-source doses are case inputs

Independent calculation instructions: Sum source weekly doses and identify dominant source.

Software results

Compared against a locked regression baseline. Comparison state: PASS · Verification state: SOFTWARE_VERIFICATION_PASS

Software result values
FieldValue
combinedWeeklyMgy0.070000
meetsGoaltrue
dominantSourceIdsrc-1

Input hash: 46e93dfcea265fa9 · Result hash: be9841b5c7703eaa

Reviewer packet

Full markdown packet a qualified reviewer uses to perform an independent calculation.

# Reviewer packet — PV-RAD-SRC-001

**Title:** Multiple source positions
**Version:** 1.0.0
**Modality:** radiography
**Software state:** SOFTWARE_VERIFICATION_PASS
**Independent review status:** READY_FOR_INDEPENDENT_REVIEW (pending human entry)

## Boundaries

- Software verification is not professional validation.
- Self-review does not qualify as independent review.
- Do not fabricate independent results.
- Licensed package citation: `ncrp147-diagnostic-uat-0.3.0` (METADATA_CITATION_ONLY).
- Numeric software lock uses demonstration transmission parameters only.

## Purpose

Aggregate multi-source weekly contributions.

## Formula IDs

- `RAD.MULTI_SOURCE.AGGREGATE`

## Independent calculation instructions

Sum source weekly doses and identify dominant source.

## Required reviewer qualification

Diagnostic medical physicist or qualified expert

## Software results (comparison values; not pre-rounded away)

```json
{
  "combinedWeeklyMgy": 0.07,
  "meetsGoal": true,
  "dominantSourceId": "src-1"
}
```

## Display rounding (informational only)

```json
{
  "combinedWeeklyMgy": "0.070000",
  "meetsGoal": "true",
  "dominantSourceId": "src-1"
}
```

## Hashes

- Input hash: `46e93dfcea265fa9`
- Result hash: `be9841b5c7703eaa`

## Assumptions

- Precomputed per-source weekly doses

## Known limitations

- Per-source doses are case inputs

Independent result entry

Demo-only entry for this CONTROLLED_UAT session. Values are held in server memory (not a database) and reset when the server restarts.

SYNTHETIC / DEMO ENTRY — not a real independent professional review

Independently computed values

Discrepancies

No discrepancies opened for this case.

Professional approval attempt

Approval is blocked while any discrepancy is open, while the case has not passed software verification, when no independent result has been recorded, or when credential/authority requirements are not met. Any approval recorded in this environment is a CONTROLLED UAT ELECTRONIC APPROVAL RECORD — not a qualified digital signature — and does not itself constitute professional approval.